Compact IO™ Isolated HART Analog Output Module
User’s Manual 0300217-03 Rev. A
7-44
Table 7-19
Tag Name Value in Hex Description
HART_PASS_THRU_QRY_RX[0] 00 HART channel
HA RT_PASS_THRU_QRY_RX[1] 00 Command Status
HA RT_PASS_THRU_QRY_RX[2]
15
Length of handle + HART response
data (Byte 1)
HA RT_PASS_THRU_QRY_RX[3]
0
Length of handle + HART response
data (Byte 2)
HART_PASS_THRU_QRY_RX[4] 02 Message handle
HA RT_PASS_THRU_QRY_RX[5] 86 Start charc ter
HART_PASS_THRU_QRY_RX[6] BE Long address byte 0
HART_PASS_THRU_QRY_RX[7] 02 Long address byte 1
HART_PASS_THRU_QRY_RX[8] 0C Long address byte 2
HART_PASS_THRU_QRY_RX[9] 77 Long address byte 3
HART_PASS_THRU_QRY_RX[10] 37 Long address byte 4
HART_PASS_THRU_QRY_RX[11] 23 HART command = 35 decimal
HART_PASS_THRU_QRY_RX[12] 0B Byte count = 11 decimal
HART_PASS_THRU_QRY_RX[13] 00 Status Byte 0
HART_PASS_THRU_QRY_RX[14] 00 Status Byte 1
HART_PASS_THRU_QRY_RX[15] 20 Range units code = 32 decimal
HA RT_PASS_THRU_QRY_RX[16] 44
HA RT_PASS_THRU_QRY_RX[17] 16
HA RT_PASS_THRU_QRY_RX[18] 00
HA RT_PASS_THRU_QRY_RX[19] 00
HA RT_PASS_THRU_QRY_RX[20] C3
HA RT_PASS_THRU_QRY_RX[21] 16
HA RT_PASS_THRU_QRY_RX[22] 00
HA RT_PASS_THRU_QRY_RX[23] 00
HART_PASS_THRU_QRY_RX[24] F9 Checksum
Upper Range value (This is a
floating point value = 600) Note: The
bytes are in reverse order.
Low er Range value (This is a
floating point value = -150) Note:
The bytes are in reverse order.